EMG-Main amplifier & interface unit
This (large) main unit supplies power to the electrodes, receives and conditions the EMG signals providing eight analog outputs.
The (small) belt-mounted interface unit is attached to the main amplifier by a cable. Each channel is labeled in correspondence with main amplifier channels and all signals are measured with respect to the reference input lead.
There are three amplification settings that provide an overall system gain of 100 (40dB), 1000 (60dB) or 10000 (80dB); these gains include the pre-amplification gain of 10 within the electrode itself. The AUX setting provides a fixed overall gain of 1000 but with a wider operating bandwidth (20 – 2000 Hz). At a gain setting of 0 the channel input is forced to be at the reference potential. When channels are not in use their gain should be set to 0.
Two LEDs indicate signal saturation (±4.8V) and line interference (60Hz). The main unit contains a tone switch which enables an audible tone when either condiiton occurs.
The power supply provides a regulated 12V output and conforms to the IEC 60601-1 safety standard.
An in-house EMG Analysis was performed in February 2004, which verified the frequency response for each channel & the gain settings for the 4-channel unit which was used at the time.
Subject Safety
[The following information was provided by Delsys.] Subject safety is
insured by the use of an isolated medical grade power supply that conforms
to the medical IEC 601 standards. The issue of electrical safety with
the power supply is addressed through various mechanisms. The many technical
points to consider are spelled out in the IEC 601-1 family of standards
for medical electrical devices.
Isolation is achieved with an isolated transformer.There are many considerations
for how to build this transformer - the windings must be isolated a certain
way, the creepage and clearances must meet minimum specifications, the
leakage current must be controlled, etc.
There is a double fuse inside the power supply, and current is limited
with specially designed regulators. This wall mounted supply has leakage
currents less than 10 uA and is safety isolated to 3750 VRMS.
The power suppply is also designed and tested to ensure that it conforms
with EMC (electro-magnetic compatibilty) requirements. Also the enclosure
must have certain mechanical characteristics and the power supply cord
must be a certain type.
The issue of subject safety is not limited to only the power supply but
is really considered for the device as a whole. For example, the Bagnoli
main amplifiers also have isolated power to protect the subject from the
output side (the power supply protects the subject from the input side.)
In this particular case, the output isolation is achieved by isolation
transformers or in more recent Delsys products by optical isolation.
|